cvedb.io
CVE-2015-3956
CRITICAL · CVSS 9.8
EPSS exploitation probability: 0%
Published 2019-03-25T18:29:00.323 · Last modified 2026-06-17T00:26:32.327

Summary

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected products

pifzer — plum_a\+_infusion_system_firmware

Does this affect you?

Add your gear to cvedb and we'll alert you only when pifzer ships something exploited.

Check my exposure →

References

This product uses data from the NVD API but is not endorsed or certified by the NVD. Informational only; not professional security advice.