CVE-2019-10964 — CVE-2019-10964

Published 2019-06-28 · Modified 2026-06-17 · Vendor medtronic · Product minimed_508_firmware · Source nvd
HIGH
severity
CVSS-derived band
7.1
CVSS v3
0–10 scale
0.0123
EPSS probability
exploitation probability, 30d
66.0%
EPSS percentile
percentile vs all CVEs
NOT LISTED
CISA KEV
known exploited catalog

Description

Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Remediation

No vendor-published fix data in our corpus for this CVE. Check the references below or the vendor's PSIRT / security advisories page.

References

cvedb.io · NVD · CISA KEV · FIRST EPSS · vendor advisories (CVE Program List v5). Informational only, no warranty — verify every remediation against the vendor advisory before acting on it. This product uses data from the NVD API but is not endorsed or certified by the NVD, CISA, FIRST.org or any vendor named. CVE® is a registered trademark of The MITRE Corporation.