CVE detail

CVE-2019-13543 — CVE-2019-13543

Published 2019-11-08 · Modified 2026-06-17 · Vendor medtronic · Product valleylab_exchange_client · Source nvd
MEDIUM
severity
CVSS-derived band
5.8
CVSS v3
0–10 scale
0.0201
EPSS probability
exploitation probability, 30d
79.0%
EPSS percentile
percentile vs all CVEs
NOT LISTED
CISA KEV
known exploited catalog

Description

Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use multiple sets of hard-coded credentials. If discovered, they can be used to read files on the device.

Remediation

No vendor-published fix data in our corpus for this CVE. Check the references below or the vendor's PSIRT / security advisories page.

References

cvedb.io · NVD · CISA KEV · FIRST EPSS · vendor advisories (CVE Program List v5). Informational only, no warranty — verify every remediation against the vendor advisory before acting on it. This product uses data from the NVD API but is not endorsed or certified by the NVD, CISA, FIRST.org or any vendor named. CVE® is a registered trademark of The MITRE Corporation.