CVE detail
CVE-2020-12040 — CVE-2020-12040
Published 2020-06-29 · Modified 2026-06-17 · Vendor baxter · Product sigma_spectrum_infusion_system_firmware · Source nvd
CRITICAL
severity
CVSS-derived band
0.0094
EPSS probability
exploitation probability, 30d
58.0%
EPSS percentile
percentile vs all CVEs
NOT LISTED
CISA KEV
known exploited catalog
Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Remediation
No vendor-published fix data in our corpus for this CVE. Check the references below or the vendor's PSIRT / security advisories page.
References