CVE detail
CVE-2020-12041 — CVE-2020-12041
Published 2020-06-29 · Modified 2026-06-17 · Vendor baxter · Product sigma_spectrum_infusion_system_firmware · Source nvd
CRITICAL
severity
CVSS-derived band
0.0140
EPSS probability
exploitation probability, 30d
70.0%
EPSS percentile
percentile vs all CVEs
NOT LISTED
CISA KEV
known exploited catalog
Description
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
Remediation
No vendor-published fix data in our corpus for this CVE. Check the references below or the vendor's PSIRT / security advisories page.