CVE detail
CVE-2020-12048 — CVE-2020-12048
Published 2020-06-29 · Modified 2026-06-17 · Vendor baxter · Product phoenix_x36_firmware · Source nvd
HIGH
severity
CVSS-derived band
0.0045
EPSS probability
exploitation probability, 30d
37.0%
EPSS percentile
percentile vs all CVEs
NOT LISTED
CISA KEV
known exploited catalog
Description
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Remediation
No vendor-published fix data in our corpus for this CVE. Check the references below or the vendor's PSIRT / security advisories page.