CVE detail
CVE-2020-25166 — CVE-2020-25166
Published 2022-04-14 · Modified 2026-06-17 · Vendor bbraun · Product datamodule_compactplus · Source nvd
HIGH
severity
CVSS-derived band
0.0046
EPSS probability
exploitation probability, 30d
38.0%
EPSS percentile
percentile vs all CVEs
NOT LISTED
CISA KEV
known exploited catalog
Description
An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.
Remediation
No vendor-published fix data in our corpus for this CVE. Check the references below or the vendor's PSIRT / security advisories page.