CVE detail

CVE-2025-14096 — CVE-2025-14096

Published 2025-12-17 · Modified 2026-06-17 · Source nvd
HIGH
severity
CVSS-derived band
8.4
CVSS v3
0–10 scale
0.0012
EPSS probability
exploitation probability, 30d
2.0%
EPSS percentile
percentile vs all CVEs
NOT LISTED
CISA KEV
known exploited catalog

Description

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided

Remediation

No vendor-published fix data in our corpus for this CVE. Check the references below or the vendor's PSIRT / security advisories page.

References

cvedb.io · NVD · CISA KEV · FIRST EPSS · vendor advisories (CVE Program List v5). Informational only, no warranty — verify every remediation against the vendor advisory before acting on it. This product uses data from the NVD API but is not endorsed or certified by the NVD, CISA, FIRST.org or any vendor named. CVE® is a registered trademark of The MITRE Corporation.